Context of Use (COU) Template — MIME#

This template helps downstream users define their Context of Use per ASME V&V 40, adapted for microrobotics simulation.

1. Question of Interest#

What does the model predict?

Example: “Predict the trajectory and drug release profile of a helical microrobot navigating from the lateral ventricle to the third ventricle via the foramen of Monro.”

2. Quantities of Interest#

Which specific output variables?

Example: robot position (x,y,z) vs. time, drug concentration at target site vs. time, step-out frequency for the specific robot geometry.

3. Model Influence#

How much does the simulation drive the decision?

Level

Description

Low

Simulation is one of several information sources

Medium

Simulation is the primary information source but clinical judgement dominates

High

Simulation output directly determines the clinical decision

4. Decision Consequence#

What is the severity of harm if the simulation is wrong?

Level

Description

Low

No injury; financial or time cost only

Medium

Non-serious injury is possible

High

Death or serious injury is possible

5. Risk Assessment#

Model influence x Decision consequence = Required evidence level.

6. Credibility Goals#

What verification, validation, and UQ evidence is needed?

  • Code verification: which MIME benchmarks (B0–B5) are relevant?

  • Calculation verification: mesh convergence, solver settings for the specific COU

  • Validation: experimental comparisons for the specific anatomy and robot design

  • Uncertainty quantification: which parameters are uncertain, what propagation method?

7. Evidence Plan#

How will the required evidence be generated?